Advanced eToims device with probe holder and electrodes

The device and electrodes are NOT available for sale in the USA. Contact us for information of availability for purchase.

Category:

Description

The front panel and controls comprise a large LCD display, push buttons and rotary controls. One of the rotary controls is solely used for changing the stimulation current whilst the other is used in conjunction with the LCD display and push buttons to present the Operator with a simple menu structure for modifying the unit’s parameters.

Using a simple control interface, the Operator has control over Stimulus Current, Pulse Width and Pulse Repetition rate. Stimulus Current and Pulse Width can also be controlled via the keypad on the ET127 Probe.

The LCD display provides parameter feed back as well as providing a simple menu interface allowing the Operator to interact with the unit. An audible device provides confirmation of changes and status feedback. The front panel contains four menu buttons and two rotary controls, one of which is dedicated to controlling the Stimulus Current, as well as a touch-proof multi way socket into which the ET127 Probe is connected.

Specifications

eToims® ET127 TECHNICAL DATA

Indications for Use
eToims® ET127 a constant current muscle stimulator. It is used for evoking muscle twitch responses at motor points (nerve-muscle meeting points commonly called trigger points) in individuals with muscle pain and/or discomfort, muscle weakness, muscle spasm or aberrant sensation. It is designed to be used by personnel trained in The eToims® Technique of motor point stimulation.

Mains voltage selector
Selects between 115 and 230 VAC voltage

Electrical characteristics
Output is completely inhibited when the unit is not timing a session.
– Maximal stimulus repetition rate is internally limited to 3 Hz (10%). Internal repetition rate adjustable between OFF and 3 Hz in 0.1 Hz steps (5%).
– Pulse widths are selectable from 50, 100, 200, 300, 400, 500 and 1000 µs.

Inputs:
– Rear panel has a 3.5 mm mono Jack Socket connector. This input has no effect when the unit is inhibited.
– Mains inlet in rear panel receives a standard IEC main plug. There is a combined power switch to enable the user to switch unit off without having to remove the power lead.

Definitions:
Credit: Total remaining time that unit will allow patient diagnosis. Credit is refilled by obtaining a specific refill code after payment in the e-commerce section of www.etoims.com.

Session Timer: Maintains the remaining time for the current diagnosis session to run. When the Session Timer has run down to zero, the Session State is changed to Stopped. The Session Timer is maintained as Minutes and Seconds.

Session State: There are three session states: “Stopped”, “Paused” and “Running”.
– Stopped: In this state, output is inhibited. Session Timer can be incremented and the menu navigated.
– Paused: Temporarily stops Session Timer with output being inhibited. Session Timer can be incremented but it is not possible to navigate the menu.
– “Running”: Output is not inhibited, stimulus pulses are produced at the rate displayed on the LCD, Session Timer counts down.

Dimensions:
Size: 265 x 110 x 290 mm – over handles (d x h x w)
Weight: 2.1 kg (approx.)

Temperature:
Operating Range: 5ºC to 30ºC
Storage Range: -40ºC to 60ºC Transport Range: -40ºC to 60ºC

Other:
Power: 100 – 120V or 200 – 240V @ 47-63 Hz
Rating: <30VA

User controls:
– Rotary control is on the front panel adjacent to the left of LCD and Stimulus current select control is adjacent to the right of the LCD.
– Stimulus current select control and pulse width controls are also available on the probe.
– Green 9-way isolated socket (Lemo Redel) type of connector is present on the extreme right of the LCD. This socket is for inserting the connector for the probe.
– Option Select (Function) Buttons on the front panel: Four grey rubber buttons equally spaced vertically aligned along the left edge of the LCD display are used in conjunction with prompts on the display to alter the stimulator parameters.

External Fuses:
Note: Should ONLY be changed by qualified persons.
100 – 120V mains T 500 mA L 5 x 20 mm, 250V Setting 115V
200 – 240V mains T 250 mA L 5 x 20 mm, 250V Setting 230V

Internal Fuses: Note: Should ONLY be changed by qualified persons.
Fuse Reference: F3 and F4 – T 500 mA L – 5 x 20 mm, 250V

 